Hach TU5300sc Low-Range Turbidimeter, LXV445.99.10112
Description
- Patented optical 360° x 90° Detection™ for maximum accuracy and repeatability
- Small sample volume delivers immediate readings and allows faster cleaning
These turbidimeters deliver outstanding low-level precision and sensitivity in drinking water and other high-purity applications. Patented optical 360° x 90° Detection™ captures 90° scatter through 360° of detection for maximum accuracy. Meters deliver immediate readings, thanks to their small sample surface chamber volume—up to 98% smaller than other meters. This also makes cleaning much easier.
Sensors track internal temperature and humidity, alerting you to condensation issues. They use quick-replacement sealed vials for easy calibration. An active drying system with desiccant cartridge and air pump circulates dried air throughout the sample cell compartment.
Turbidimeter includes: mounting bracket (fits pattern for legacy 1720E sensors), service vial removal tool, desiccant cartridge, flow regulator valve, integral 5'L digital cable, and remote-mounted maintenance bracket.
Optional accessories are strongly recommended to ensure reliable results. In-line flow sensor (stock # 93652) lets you display inlet flow rate. Automatic cleaning module (stock # 93651) uses a motorized wiper that's set to user-defined intervals (replaces the standard chamber lid).
Note: Turbidimeter must be used in conjunction with a Hach digital controller, such as an SC4500 or SC1000 (sold separately), or legacy SC200. Calibration lid is also required; purchase separately (stock # 93661) or with maintenance kit (stock # 72855). If using Prognosys™ diagnostics system on an sc1000 controller, you must order turbimeter with internal system check capabilities (stock # 93648 or 93650).
- Range: 0 to 700 NTU, FNU & FTU
- Accuracy: ±2% of reading plus 0.01 NTU (up to 40 NTU), ±10% of reading (over 40 NTU); based on Formazin at 25°C (77°F).
- Repeatability: <1% of reading or ±0.002 NTU, whichever is greater
- Response Time: T90 <30 seconds at 100 mL/min
- Sample flow rate: 100 to 1000 mL/min (1.6 to 15.9 gal/hour); optimal: 200 to 500 mL/min (3.2 to 7.9 gal/hour)
Product Information
- Item number
- 93647
- Brand Name
- Hach Company
- Catalog Page
- 1136
- Net weight
- 13
- Condition
- New
Accessories
Related
Hach TU5300sc Low-Range Turbidimeter, LXV445.99.10112
- Item number
- 93647
- Brand Name
- Hach Company
- Catalog Page
- 1136
- Net weight
- 13
- Condition
- New
- Patented optical 360° x 90° Detection™ for maximum accuracy and repeatability
- Small sample volume delivers immediate readings and allows faster cleaning
These turbidimeters deliver outstanding low-level precision and sensitivity in drinking water and other high-purity applications. Patented optical 360° x 90° Detection™ captures 90° scatter through 360° of detection for maximum accuracy. Meters deliver immediate readings, thanks to their small sample surface chamber volume—up to 98% smaller than other meters. This also makes cleaning much easier.
Sensors track internal temperature and humidity, alerting you to condensation issues. They use quick-replacement sealed vials for easy calibration. An active drying system with desiccant cartridge and air pump circulates dried air throughout the sample cell compartment.
Turbidimeter includes: mounting bracket (fits pattern for legacy 1720E sensors), service vial removal tool, desiccant cartridge, flow regulator valve, integral 5'L digital cable, and remote-mounted maintenance bracket.
Optional accessories are strongly recommended to ensure reliable results. In-line flow sensor (stock # 93652) lets you display inlet flow rate. Automatic cleaning module (stock # 93651) uses a motorized wiper that's set to user-defined intervals (replaces the standard chamber lid).
Note: Turbidimeter must be used in conjunction with a Hach digital controller, such as an SC4500 or SC1000 (sold separately), or legacy SC200. Calibration lid is also required; purchase separately (stock # 93661) or with maintenance kit (stock # 72855). If using Prognosys™ diagnostics system on an sc1000 controller, you must order turbimeter with internal system check capabilities (stock # 93648 or 93650).
- Range: 0 to 700 NTU, FNU & FTU
- Accuracy: ±2% of reading plus 0.01 NTU (up to 40 NTU), ±10% of reading (over 40 NTU); based on Formazin at 25°C (77°F).
- Repeatability: <1% of reading or ±0.002 NTU, whichever is greater
- Response Time: T90 <30 seconds at 100 mL/min
- Sample flow rate: 100 to 1000 mL/min (1.6 to 15.9 gal/hour); optimal: 200 to 500 mL/min (3.2 to 7.9 gal/hour)