HF scientific MCX Monochloramine Analyzer, Single Sample Stream Inlet, 28056

Item No.403165
Product Details 3-in-1 system displays monochloramine, total ammonia and free ammonia Provides accurate readings every 20 minutes Continuous operation for 30 days between reagent changeouts Can be expanded to measure a second sample stream (optional sample stream sequencer required) The MCX Monochlo
Call for availability
Call for pricing & to order

Description

Product Details
  • 3-in-1 system displays monochloramine, total ammonia and free ammonia
  • Provides accurate readings every 20 minutes
  • Continuous operation for 30 days between reagent changeouts
  • Can be expanded to measure a second sample stream (optional sample stream sequencer required)

The MCX Monochloramine Analyzer measures, logs and displays the concentration of monochloramine and ammonia in water samples. Ideal for water plants that use chloramination for disinfection, it uses the modified phenate method to test for monochloramine and total ammonia, and calculates free ammonia.

The analyzer takes samples of the process stream of water at user-adjustable intervals and shows readings on its large display. Readings are also accessible remotely via built-in Modbus RS-485 communication and 4-20 mA outputs (one each for monochloramine, total ammonia and free ammonia). Two user-programmable relay outputs let you select one of four alarm states: error, warning, end of cycle or reaction complete. A micro-USB port allows you to download readings and error logs.

The MCX analyzer requires a set of three reagents (sold separately; stock # 403137) for operation. Reagents typically last for 30 days of continuous operation for easy maintenance.

All operating components and reagents are housed within a metallic NEMA 4X enclosure panel. The enclosure features four mounting tabs for wall mounting. An indoor location is recommended. Unit features cable glands for electrical connections, and two simple plumbing push-to-connect fittings for the sample inlet and drain outlet.

Note: Due to variability in applications, please contact our Technical Specialists prior to purchase. They will ensure all necessary application details are covered, including the potential need for on-site factory commissioning and training.

For a complete system, order: MCX Monochloramine Analyzer, Reagent Kit (stock # 403137) and Optional Accessories as required by your application.

Tech Specs:
  • Display: 5" color graphic, 7 pushbuttons
  • Measurement range, monochloramine: 0 to 10 mg/L (as NH2Cl-Cl2) or 0.02 to 2.0 mg/L (as NH2Cl-N)
  • Measurement range, total ammonia: 0.02 to 2 mg/L (as NH3-N)
  • Measurement interval: 20 minutes to 24 hours (adjustable)
  • Accuracy: ±5% or 0.01 ppm (as N) for 61°F to 104°F; ±10% or 0.02 ppm for 104°F to 122°F, whichever is greater
  • Sample pressure: 18 psi max, 122°F max
  • Sample flow rate: 185 mL/min to 225 mL/min (0.044 to 0.059 gal/min)
  • Connections: push-to-connect fittings, 1/4" inlet, 3/8" drain
  • Communications: Modbus RS485
  • Data logger: readings, data and errors; micro-USB port data retrieval
  • Analog outputs: three 4-20 mA, isolated
  • Relays: two SPDT, unpowered
  • Power: 100-240 VAC, 47/63 Hz
  • Enclosure: metal, NEMA 4X
  • Dimensions: 20.2"W x 24.5"H x 9.9"D

Product Information

Item number
403165
Brand Name
HF Scientific
Net weight
83.6
Condition
New

HF scientific MCX Monochloramine Analyzer, Single Sample Stream Inlet, 28056

Item number
403165
Brand Name
HF Scientific
Net weight
83.6
Condition
New
Call for availability
Call for pricing & to order
Need Help? Call 800-548-1234
Product Details
  • 3-in-1 system displays monochloramine, total ammonia and free ammonia
  • Provides accurate readings every 20 minutes
  • Continuous operation for 30 days between reagent changeouts
  • Can be expanded to measure a second sample stream (optional sample stream sequencer required)

The MCX Monochloramine Analyzer measures, logs and displays the concentration of monochloramine and ammonia in water samples. Ideal for water plants that use chloramination for disinfection, it uses the modified phenate method to test for monochloramine and total ammonia, and calculates free ammonia.

The analyzer takes samples of the process stream of water at user-adjustable intervals and shows readings on its large display. Readings are also accessible remotely via built-in Modbus RS-485 communication and 4-20 mA outputs (one each for monochloramine, total ammonia and free ammonia). Two user-programmable relay outputs let you select one of four alarm states: error, warning, end of cycle or reaction complete. A micro-USB port allows you to download readings and error logs.

The MCX analyzer requires a set of three reagents (sold separately; stock # 403137) for operation. Reagents typically last for 30 days of continuous operation for easy maintenance.

All operating components and reagents are housed within a metallic NEMA 4X enclosure panel. The enclosure features four mounting tabs for wall mounting. An indoor location is recommended. Unit features cable glands for electrical connections, and two simple plumbing push-to-connect fittings for the sample inlet and drain outlet.

Note: Due to variability in applications, please contact our Technical Specialists prior to purchase. They will ensure all necessary application details are covered, including the potential need for on-site factory commissioning and training.

For a complete system, order: MCX Monochloramine Analyzer, Reagent Kit (stock # 403137) and Optional Accessories as required by your application.

Tech Specs:
  • Display: 5" color graphic, 7 pushbuttons
  • Measurement range, monochloramine: 0 to 10 mg/L (as NH2Cl-Cl2) or 0.02 to 2.0 mg/L (as NH2Cl-N)
  • Measurement range, total ammonia: 0.02 to 2 mg/L (as NH3-N)
  • Measurement interval: 20 minutes to 24 hours (adjustable)
  • Accuracy: ±5% or 0.01 ppm (as N) for 61°F to 104°F; ±10% or 0.02 ppm for 104°F to 122°F, whichever is greater
  • Sample pressure: 18 psi max, 122°F max
  • Sample flow rate: 185 mL/min to 225 mL/min (0.044 to 0.059 gal/min)
  • Connections: push-to-connect fittings, 1/4" inlet, 3/8" drain
  • Communications: Modbus RS485
  • Data logger: readings, data and errors; micro-USB port data retrieval
  • Analog outputs: three 4-20 mA, isolated
  • Relays: two SPDT, unpowered
  • Power: 100-240 VAC, 47/63 Hz
  • Enclosure: metal, NEMA 4X
  • Dimensions: 20.2"W x 24.5"H x 9.9"D